Advanced Sterilization Products STERRAD NX Sterilizer, Product Code 10033 Low temperature sterilizers that use hydrogen peroxide gas plasma to inactivate microorganisms on a broad range of medical and surgical instruments. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
STERRAD NX Sterilizer, Product Code 10033 Low temperature sterilizers that use hydrogen peroxide gas plasma to inactivate microorganisms on a broad range of medical and surgical instruments.
Brand
Advanced Sterilization Products
Lot Codes / Batch Numbers
Serial Numbers: 0033081006, 0033081007, 0033081016, 0033081017, 0033081019, 0033081020, 0033081022- 0033081029, 0033081036, 0033081039, 0033081048, 0033081059, 0033081080, 0033081082, 0033081086, 0033081088, 0033081089, 0033081091, 0033081095, 10033060892, 10033060379, 100330050671, 10033070576, 10033080067, 10033050663, and 10033050661.
Products Sold
Serial Numbers: 0033081006, 0033081007, 0033081016, 0033081017, 0033081019, 0033081020, 0033081022- 0033081029, 0033081036, 0033081039, 0033081048, 0033081059, 0033081080, 0033081082, 0033081086, 0033081088, 0033081089, 0033081091, 0033081095, 10033060892, 10033060379, 100330050671, 10033070576,10033080067,10033050663, and 10033050661.
Advanced Sterilization Products is recalling STERRAD NX Sterilizer, Product Code 10033 Low temperature sterilizers that use hydrogen peroxide ga due to ASP has discovered a component defect in some of the UV lamp power supplies used in certain STERRAD NX Sterilizers. This defect can potentially cause. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
ASP has discovered a component defect in some of the UV lamp power supplies used in certain STERRAD NX Sterilizers. This defect can potentially cause the Hydrogen Peroxide Monitor to give inaccurate readings.
Recommended Action
Per FDA guidance
The Urgent: Product Correction customer notifications, dated October 6, 2008, were hand delivered by field representatives starting October 7, 2008. The letter advises that Advanced Sterilization Products has discovered a component defect in some of the UV lamp power supplies used in certain STERRAD Sterilizers. This defect can potentially cause the Hydrogen Peroxide Monitor to give inaccurate readings. A list of the affected sterilizer serial numbers is included with the letter. The letter advises that the Field Service Engineer has come to their facility to inspect and, if necessary, replace this power supply. Customers are asked to sign the letter acknowledging receipt of this letter upon completion of the work by the Field Service Engineer.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AL, AZ, CA, DE, FL, GA, MA, MI, MO, NJ, NM, NC, SC, TN, WA, WV, WI