Airgas Puritan Medical Puritan Medical Unit AGI-2238 Portable 4 Cylinder Manifold Cart, for connecting medical gas cylinders. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
AGI-2238 Portable 4 Cylinder Manifold Cart, for connecting medical gas cylinders.
Brand
Airgas Puritan Medical Puritan Medical Unit
Lot Codes / Batch Numbers
All units of model numbers 8065-870 and 8066-910
Products Sold
All units of model numbers 8065-870 and 8066-910
Airgas Puritan Medical Puritan Medical Unit is recalling AGI-2238 Portable 4 Cylinder Manifold Cart, for connecting medical gas cylinders. due to Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.
Recommended Action
Per FDA guidance
Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026