Apogee Medical, Inc Apogee Intermittent Catheter, 8 Fr 16", Catalog/Ref No. 1018, Sterile, Latex free. The device is intended to be used as a urinary incontinence device designed to drain urine from the bladder. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Apogee Intermittent Catheter, 8 Fr 16", Catalog/Ref No. 1018, Sterile, Latex free. The device is intended to be used as a urinary incontinence device designed to drain urine from the bladder.
Brand
Apogee Medical, Inc
Lot Codes / Batch Numbers
Lot number 7206
Products Sold
Lot number 7206
Apogee Medical, Inc is recalling Apogee Intermittent Catheter, 8 Fr 16", Catalog/Ref No. 1018, Sterile, Latex free. The device is in due to Due to a package seal defect the catheter tip may have been inadvertently cut which could cause trauma to the urethra upon use. Also the sterility of . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to a package seal defect the catheter tip may have been inadvertently cut which could cause trauma to the urethra upon use. Also the sterility of the product could be compromised.
Recommended Action
Per FDA guidance
Distributors and Retailers that have received any of the Catalog Numbers listed in the letter titled URGENT MEDICAL DEVICE RECALL were notified on/about July 17, 2008. Recipients of the letter are being requested to review their inventory and return any of the applicable Catalog Items from the specific lot numbers listed in the recall letter. Distributors are being instructed to notify their customers and inform them of the recall with a copy of the recall letter. Their notification should instruct their customers to return the recalled product to them so that they could return the product to Apogee Medical LLC. In addition to the initial letter, follow-up letters, telephone calls and emails will be utilized to assess the effectiveness of the recall. A spreadsheet has been set up and will be maintained in order to easily identify the customers that have been contacted, those who have responded and those who have returned product. If you have questions, contact Diane Peper at 919-435-5409.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026