Arjo, Inc. BHM Medical Scales, 1000 lb. scales weight limit; Part Number 700.00525; Manufactured by BHM Medical Inc., Magog, QC, Canada; These scales were sold for installation on the Model V10 and Model Maxisky fixed ceiling lifts manufactured by BHM Medical. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BHM Medical Scales, 1000 lb. scales weight limit; Part Number 700.00525; Manufactured by BHM Medical Inc., Magog, QC, Canada; These scales were sold for installation on the Model V10 and Model Maxisky fixed ceiling lifts manufactured by BHM Medical.
Brand
Arjo, Inc.
Lot Codes / Batch Numbers
Part 700.00525, Serial Numbers: ERSC-3064, ERSC-3065, ERSC-3066, ERSC-3068, ERSC-5858, ERSC-5859, ERSC-5860, ERSC-5861, ERSC-5864, ERSC-5866, ERSC-6765, ERSC-6767, ERSC-6768, ERSC-7729, ERSC-7742, ERSC-7769, ERSC-7776, ERSC-7779, ERSC-7780, ERSC-7781, ERSC-7784, ERSC-7787, ERSC-7789, ERSC-7790, ERSC-7791, and ERSC-8038
Products Sold
Part 700.00525, Serial Numbers: ERSC-3064, ERSC-3065, ERSC-3066, ERSC-3068, ERSC-5858, ERSC-5859, ERSC-5860, ERSC-5861, ERSC-5864, ERSC-5866, ERSC-6765, ERSC-6767, ERSC-6768, ERSC-7729, ERSC-7742, ERSC-7769, ERSC-7776, ERSC-7779, ERSC-7780, ERSC-7781, ERSC-7784, ERSC-7787, ERSC-7789, ERSC-7790, ERSC-7791, and ERSC-8038,
Arjo, Inc. is recalling BHM Medical Scales, 1000 lb. scales weight limit; Part Number 700.00525; Manufactured by BHM Medical due to Unintended Hanger Bar Detachment: The locknut on the scales may loosen during use and allow the hanger bar assembly of the patient lift to detach and. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Unintended Hanger Bar Detachment: The locknut on the scales may loosen during use and allow the hanger bar assembly of the patient lift to detach and fall.
Recommended Action
Per FDA guidance
Arjo sent Urgent Device Field Correction Notification letters dated 12/19/07 to the end user accounts who received the affected scales, advising them of the potential for the hanger bar assembly to detach. The accounts were given the option immediately stopping use of the lifts with the affected scales or removing the scales from the lift so the lift can continue to be used. Instructions for removal of the scales were provided with the letter. The accounts were requested to complete and return the enclosed customer response form, acknowledging the receipt and understanding of the letter, and indicating whether they will complete the correction upon receipt of the correction parts or if they require the assistance of an Arjo Service Technician.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026