Arjo, Inc. Medi-Lifter III Plus Total Lift, Model 7000-2000 and Summit Total Lift, Model 4800-2000; patient lifts with mast scale (600 lbs. capacity); Mast scale is Intended to be used for weighing patients while in the lifter. Manufactured by Medi-Man Rehabilitation Products Inc., 6200A Tomken Road, Mississauga, Ontario, Canada L5T 1X7, now owned by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Medi-Lifter III Plus Total Lift, Model 7000-2000 and Summit Total Lift, Model 4800-2000; patient lifts with mast scale (600 lbs. capacity); Mast scale is Intended to be used for weighing patients while in the lifter. Manufactured by Medi-Man Rehabilitation Products Inc., 6200A Tomken Road, Mississauga, Ontario, Canada L5T 1X7, now owned by BMH Medical Inc., 2001 Tanguay Street , Magog, Quebec, Canada J1X 5Y5
Brand
Arjo, Inc.
Lot Codes / Batch Numbers
Medi-Lifter III Plus, serial numbers 76001/1013, 76003/1015, 76004/1014, 76005/1017, 76027/1019, 76028/1012, 76058/1024, 76128/1034, 76129/1025, 76132/1031, 76133/1039, 76162/1033, 76164/1035, 76221/1069, 76247/1065, 76251/1070, 77091/1079, 77092/1088, 77093/1083, 77126/1084, 77127/1085, 77295/1091, 77342/1096, 77343/1093, 77406/1099, 77407/1098, 77408/1092, 78035/1134, 78036/1132, 78037/1133, 78038/1128, 78039/1103, 78040/1130, 78041/1104, Summit, serial numbers 4800157/1020, 4800166/1077, 4800173/1086
Products Sold
Medi-Lifter III Plus, serial numbers 76001/1013, 76003/1015, 76004/1014, 76005/1017, 76027/1019, 76028/1012, 76058/1024, 76128/1034, 76129/1025, 76132/1031, 76133/1039, 76162/1033, 76164/1035, 76221/1069, 76247/1065, 76251/1070, 77091/1079, 77092/1088, 77093/1083, 77126/1084, 77127/1085, 77295/1091, 77342/1096, 77343/1093, 77406/1099, 77407/1098, 77408/1092, 78035/1134, 78036/1132, 78037/1133, 78038/1128, 78039/1103, 78040/1130, 78041/1104; Summit, serial numbers 4800157/1020, 4800166/1077, 4800173/1086
Arjo, Inc. is recalling Medi-Lifter III Plus Total Lift, Model 7000-2000 and Summit Total Lift, Model 4800-2000; patient lif due to Scale could break and fall potentially causing injuries to someone who is on it. There is a potential for the mast scale load cell ssembly to fractur. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Scale could break and fall potentially causing injuries to someone who is on it. There is a potential for the mast scale load cell ssembly to fracture and allow the mast/boom to fall. Should a patient be in the lift at the time, the patient could fall and sustain injuries.
Recommended Action
Per FDA guidance
Arjo sent Urgent Device Recall Customer Notification letters dated 4/1/08 on the BHM Medical letterhead to the end user accounts who received the affected patient lifts with mast scales, advising them of the potential of fracture at the load cell assembly which might cause the scale to fall. Should a patient be in the lift at the time, the patient could fall, potentially sustaining injuries. The accounts were instructed to immediately stop using the patient lifts. A BHM representative will contact the accounts within 5 days to arrange for the correction of the lifts. The accounts will be given the option of returning the lifts for repair (with shipping costs covered by BHM), or purchasing a replacement lift at a significantly reduced price. Any questions or concerns were directed to the BHM Service Department at 1-819-868-0441 or service@bhm-medical.com.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026