Arrow International, Inc. 8 Fr 30cc Ultra 8 IAB, IAB-05830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the aorta during diastole and deflation during systole increase blood supply to the heart muscle and decrease work of the left ventricle. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
8 Fr 30cc Ultra 8 IAB, IAB-05830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the aorta during diastole and deflation during systole increase blood supply to the heart muscle and decrease work of the left ventricle.
Brand
Arrow International, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Lot/Serial Numbers: MF6122058 MF7065640 MF7076257 MF7076378 MF7097392 MF7118003 MF7118004 MF7118219 MF7128387 MF7128471 MF7128504 MF7128505 MF8018907 MF8018997 MF8030122 MF8039664 MF8039878 MF8040324 MF8040325 MF8040638 MF8050911 MF8061125 MF8061176 MF8061334 MF8061500 MF8061615 MF8071782 MF8072074 MF8082267 MF8082584 MF8092783 MF8092784 MF8093000 MF8093001 MF8103155 MF8103262 MF8103263 MF8124064 MF8124178 MF8124064 MF8124307 and MF8124408.
Arrow International, Inc. is recalling 8 Fr 30cc Ultra 8 IAB, IAB-05830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) i due to A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP mo. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Recommended Action
Per FDA guidance
An "Urgent Medical Device Recall" letter dated February 2, 2009 was issued to consignees via US Certified Mail to notify users of the problem. Users were instructed to check their inventory for the affected products, quarantine, and return the them to Arrow International, Inc. Customers were also instructed to follow the "Mitigation Directions" in circumstances where alternate Intra-Aortic Balloon (IAB) catheters/adaptors or other forms of therapy are not available (until receipt of available product). Canadian accounts were notified on February 9, 2009 by courier and International accounts notified on 6 February 2009 and 10 February 2009 via Registered Post/UPS. Direct questions to your local Sales Representative or the Arrow International Customer Service at 1-800-523-8446 (US) or 1-919-361-4062 (outside US). Press Release issued April 13, 2009 is posted to firm's webiste: http://www.arrowintl.com/iab_recall.asp.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.