Arrow International, Inc. 8 Fr 40cc Narrow Flex IAB Catheter Kit with Generic Adapters Product Number: IAB-04840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the aorta during diastole and deflation during systole increase blood supply to the heart muscle and decrease work of the left ventricle. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
8 Fr 40cc Narrow Flex IAB Catheter Kit with Generic Adapters Product Number: IAB-04840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the aorta during diastole and deflation during systole increase blood supply to the heart muscle and decrease work of the left ventricle.
Brand
Arrow International, Inc.
Lot Codes / Batch Numbers
Not specified in FDA notice. Check official source for details.
Products Sold
Lot/ASerial Numbers: MF6090318 MF6111511 MF7065955 MF7076255 MF7086556 MF7097114 MF7097115 MF7097226 MF7097227 MF7097228 MF7097427 MF7097470 MF7097471 MF7107861 MF7107862 MF7118045 MF7118046 MF7128470 MF7128508 MF8018724 MF8018813 MF8018996 MF8124303 MF8124304 MF8124304 MF9014426 MF9014444 MF9014446 and MF9014447.
Arrow International, Inc. is recalling 8 Fr 40cc Narrow Flex IAB Catheter Kit with Generic Adapters Product Number: IAB-04840-U, Intra-aort due to A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP mo. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.
Recommended Action
Per FDA guidance
An "Urgent Medical Device Recall" letter dated February 2, 2009 was issued to consignees via US Certified Mail to notify users of the problem. Users were instructed to check their inventory for the affected products, quarantine, and return the them to Arrow International, Inc. Customers were also instructed to follow the "Mitigation Directions" in circumstances where alternate Intra-Aortic Balloon (IAB) catheters/adaptors or other forms of therapy are not available (until receipt of available product). Canadian accounts were notified on February 9, 2009 by courier and International accounts notified on 6 February 2009 and 10 February 2009 via Registered Post/UPS. Direct questions to your local Sales Representative or the Arrow International Customer Service at 1-800-523-8446 (US) or 1-919-361-4062 (outside US). Press Release issued April 13, 2009 is posted to firm's webiste: http://www.arrowintl.com/iab_recall.asp.