Ascension Orthopedics, Inc Ascension PyroCarbon PIP Total Joint - Trials; Ascension PIP, sizes 10P, 20P, 30P and 40P. Ascension Orthopedics, 8700 Cameron Road, Suite 100, Austin, Texas 78754; TEL: 512 836-5001. Intended as an accessory to the Ascension PyroCarbon PIP Total Joint used in finger joint implant procedures where the joint has become painful, stiff or cannot move because of arthritis. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Ascension PyroCarbon PIP Total Joint - Trials; Ascension PIP, sizes 10P, 20P, 30P and 40P. Ascension Orthopedics, 8700 Cameron Road, Suite 100, Austin, Texas 78754; TEL: 512 836-5001. Intended as an accessory to the Ascension PyroCarbon PIP Total Joint used in finger joint implant procedures where the joint has become painful, stiff or cannot move because of arthritis.
Brand
Ascension Orthopedics, Inc
Lot Codes / Batch Numbers
All lot codes.
Products Sold
All lot codes.
Ascension Orthopedics, Inc is recalling Ascension PyroCarbon PIP Total Joint - Trials; Ascension PIP, sizes 10P, 20P, 30P and 40P. Ascensi due to Trial product was larger than actual implant. Trial product was re-designed to more closely match the implant.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Trial product was larger than actual implant. Trial product was re-designed to more closely match the implant.
Recommended Action
Per FDA guidance
Ascension Orthopedics, Inc. issued e-mail notification to distributors and sales representatives beginning June 2009. Consignees were instructed return affected product to the firm for replacement. For further information, contact Ascension Orthopedics, Inc. at 1-512-836-5001 extension 1549.