Hot Dog Patient Warming Mattress (Augustine) – Cable Connection Instability (2024)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia
Brand
Augustine Temperature Management, LLC
Lot Codes / Batch Numbers
UDI-DI UDI-DI 00855913001817, Serial Numbers: 1304292, 1304293, 1304296, 1304297
Products Sold
UDI-DI UDI-DI 00855913001817, Serial Numbers: 1304292, 1304293, 1304296, 1304297
Augustine Temperature Management, LLC is recalling Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to p due to There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, t. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Recommended Action
Per FDA guidance
Augustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026