B. Braun Medical, Inc. 8713051U, Infusomat Large Volume Pump, Wireless; Volumetric Infusion Pump System Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
8713051U, Infusomat Large Volume Pump, Wireless; Volumetric Infusion Pump System
Brand
B. Braun Medical, Inc.
Lot Codes / Batch Numbers
UDI-DI: 04046964660887 Serial Numbers: 868497 -892669 ***Updated 8/7/24*** SNs added: 430056, 430119, 430155, 430243, 551518, 685947, 703196, 705968, 726936, 737729, 753364, 774163, 790843, 814074
Products Sold
UDI-DI: 04046964660887 Serial Numbers: 868497 -892669 ***Updated 8/7/24*** SNs added: 430056, 430119, 430155, 430243, 551518, 685947, 703196, 705968, 726936, 737729, 753364, 774163, 790843, 814074
B. Braun Medical, Inc. is recalling 8713051U, Infusomat Large Volume Pump, Wireless; Volumetric Infusion Pump System due to Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications m. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.
Recommended Action
Per FDA guidance
On September 21, 2023, the firm began notifying customers of the issue via "URGENT MEDICAL DEVICE CORRECTION" letters. BBMI will schedule service to update impacted devices. Customers were provided with stickers to identify affected units while waiting for the service to be completed. In the meantime, customers were informed of measures that may aid in continued use of the pumps, such as only using non-impacted devices for the delivery of high-risk medications and placing the impacted units in care areas that do not administer high risk medications. Customers may contact BBMI by calling 1-800-627-7867 (1-800-627-PUMP) for technical support for potential alternate interim measures. ***Updated 8/7/24*** On August 7, 2024, firm identified additional units which were serviced using potentially impacted sensors. Serial numbers were added to the two models already in scope (Article Numbers 8713051U and 8713052U). Serial numbers from an additional model (Article No. 8713050U) also added and tracked under RES 95093.