Hemodialysis Bloodlines (B Braun) – micro-air bubble risk (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version: N/A Product Description: STREAMLINE BLOODLINE SET FOR DIALOG Component: N/A
Brand
B BRAUN MEDICAL INC
Lot Codes / Batch Numbers
Model No. SL-2010M2096, UDI-DI 04046964367786, Lot Numbers: A2500119, A2500121, A2500128, A2500129, A2500130, A2500132, A2500133, A2500134, A2500149, A2500150, A2500151, A2500180, A2500181, A2500182, A2500185, A2500198, A2500199, A2500200, A2500203, A2500204, A2500205, A2500206, A2500207, A2500208, A2500209, A2500210, A2500211, A2500212, A2500213, A2500214, A2500215, A2500216, A2500217, A2500218, A2500219, A2500220, A2500246, A2500247, A2500248, A2500249, A2500250, A2500251, A2500257, A2500258, A2500259, A2500260, A2500261, A2500262, A2500267, A2500268, A2500269, A2500273, A2500274, A2500275, A2500276, A2500277, A2500278, A2500295, A2500298, A2500299, A2500300
Products Sold
Model No. SL-2010M2096; UDI-DI 04046964367786; Lot Numbers: A2500119, A2500121, A2500128, A2500129, A2500130, A2500132, A2500133, A2500134, A2500149, A2500150, A2500151, A2500180, A2500181, A2500182, A2500185, A2500198, A2500199, A2500200, A2500203, A2500204, A2500205, A2500206, A2500207, A2500208, A2500209, A2500210, A2500211, A2500212, A2500213, A2500214, A2500215, A2500216, A2500217, A2500218, A2500219, A2500220, A2500246, A2500247, A2500248, A2500249, A2500250, A2500251, A2500257, A2500258, A2500259, A2500260, A2500261, A2500262, A2500267, A2500268, A2500269, A2500273, A2500274, A2500275, A2500276, A2500277, A2500278, A2500295, A2500298, A2500299, A2500300;
B BRAUN MEDICAL INC is recalling Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalo due to The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.
Recommended Action
Per FDA guidance
On September 29, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters dated September 24, 2025, were sent to customers. Actions to be taken: 1. Ensure full adherence to the product Instructions for Use (IFU). In particular, the instructions for secure connection of patient connectors should be followed to reduce the probability of air ingress or leakage. 2. Increase visual inspection of the bloodlines during treatment. If air ingress or leakage is observed, re-secure or tighten connection using the product IFU. In some instances, the dialysis treatment should be stopped. 3. If re-securement or tightening connections does not resolve the concern, replace bloodline at earliest clinically feasible time. 4. If possible, avoid high flow during treatment if microbubbles are present. B. Braun Medical Inc. (BBMI) Customer/User: 1. Review this notice in its entirety. Ensure that all users in your organization of the above-mentioned products, and other concerned personnel are informed about this voluntary correction. Post this notification where the affected products are stored. Note: Product is not being returned as part of this correction notification. a. If you are a distributor and have further distributed the product, please forward this notice to your consignees. The correction is to be extended to the consumer level. 2. Return the completed Urgent Medical Device Correction Acknowledgement Form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to recalls@bbraunusa.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). If you have any questions, contact Director, Corporate Communications at 484-240-8373.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026