Bard Peripheral Vascular Inc SAFE-T PLUS LUMBAR PUNCTURE TRAY ADULT Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SAFE-T PLUS LUMBAR PUNCTURE TRAY ADULT
Brand
Bard Peripheral Vascular Inc
Lot Codes / Batch Numbers
Catalog Number: 4301CSP Lot Numbers/UDI (GTIN
Products Sold
Catalog Number: 4301CSP Lot Numbers/UDI (GTIN, DI + PI): 0001411297 (01)10885403107771(17)220331(10)0001411297 0001412994 (01)10885403107771(17)220331(10)0001412994 0001412993 (01)10885403107771(17)220331(10)0001412993 0001413312 (01)10885403107771(17)220331(10)0001413312 0001413224 (01)10885403107771(17)220331(10)0001413224 0001415321 (01)10885403107771(17)220430(10)0001415321 0001416350 (01)10885403107771(17)220331(10)0001416350 0001416833 (01)10885403107771(17)220430(10)0001416833 0001417546 (01)10885403107771(17)220430(10)0001417546 0001418031 (01)10885403107771(17)220430(10)0001418031 0001418108 (01)10885403107771(17)220430(10)0001418108 0001418415 (01)10885403107771(17)220430(10)0001418415 0001420365 (01)10885403107771(17)220430(10)0001420365 0001421922 (01)10885403107771(17)220630(10)0001421922 0001422603 (01)10885403107771(17)220630(10)0001422603 0001422916 (01)10885403107771(17)220630(10)0001422916 0001423132 (01)10885403107771(17)220630(10)0001423132 0001424769 (01)10885403107771(17)220630(10)0001424769 0001425084 (01)10885403107771(17)220630(10)0001425084 0001426503 (01)10885403107771(17)220630(10)0001426503 0001426326 (01)10885403107771(17)220630(10)0001426326 0001427017 (01)10885403107771(17)220630(10)0001427017 0001427227 (01)10885403107771(17)220630(10)0001427227 0001427711 (01)10885403107771(17)220630(10)0001427711 0001428496 (01)10885403107771(17)220630(10)0001428496 0001429509 (01)10885403107771(17)220630(10)0001429509 0001430610 (01)10885403107771(17)220630(10)0001430610 0001430605 (01)10885403107771(17)220630(10)0001430605 0001431471 (01)10885403107771(17)220630(10)0001431471 0001432426 (01)10885403107771(17)220831(10)0001432426 0001432425 (01)10885403107771(17)220831(10)0001432425 0001433266 (01)10885403107771(17)220831(10)0001433266 0001434250 (01)10885403107771(17)220930(10)0001434250 0001437530 (01)10885403107771(17)221031(10)0001437530
Bard Peripheral Vascular Inc is recalling SAFE-T PLUS LUMBAR PUNCTURE TRAY ADULT due to Due to lumbar puncture tray manometer failure mode of leaking and mating issues.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Due to lumbar puncture tray manometer failure mode of leaking and mating issues.
Recommended Action
Per FDA guidance
Between 01/28 and 02/01/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via email and FedEx 2-day mailings to customers informing them that between 07/02 to 12/15/2021 the firm received 59 customer complaints for lumbar puncture tray manometers with reported failure modes of leaking and mating issues. Customers are asked to take the following actions: 1. Check all inventory locations within their institution for the product listed in Table A and immediately destroy all affected product remaining in their possession. The product should have the sterile barrier broken and destroyed per their local facility s destruction process. 2. Share this notification with all users within their facilities to ensure they are also aware of this Urgent Medical Device Correction. 3. If Customers purchased this product from a distributor, contact their distributor for further instructions and credit resolution. 4. Complete and return the attached Customer Response Form to the Recalling Firm contact noted on the form confirming acknowledgement of the correction notification, whether or not they have any affected product, so that the Recalling Firm may acknowledge their receipt of this notification and process their credit. 5. Indicate on the response form the quantity of affected product identified at their facility and confirm that this inventory was destroyed. 6. Contact their Recalling Firm's representative to assist in this process or they may also contact the listed in the notification for further assistance. For further assistance contact: North American Regional Complaint Center - 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted M-F 8am - 5pm CT
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026