Baxa Corporation Exacta-Mix 600 Pharmacy Compounding System is used in the pharmacy to compound multiple source ingredients into on final solution based on user-specified sequences and volumes. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Exacta-Mix 600 Pharmacy Compounding System is used in the pharmacy to compound multiple source ingredients into on final solution based on user-specified sequences and volumes.
Brand
Baxa Corporation
Lot Codes / Batch Numbers
Model 071B-RS232, all serial numbers
Products Sold
Model 071B-RS232, all serial numbers
Baxa Corporation is recalling Exacta-Mix 600 Pharmacy Compounding System is used in the pharmacy to compound multiple source ingre due to Flush solution from TPN compounder may be added to patient's TPN bag.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Flush solution from TPN compounder may be added to patient's TPN bag.
Recommended Action
Per FDA guidance
Consignees were notified by letter (Exacta-Mix Compounder Software Safety Alert) on 02/11/2008 advising them of the necessity to install a waste bag prior to a machine flush.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026