Baxter Healthcare Corp. Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication therapies requiring continuous infusion. The infusion pumps can be utilized for slow, continuous delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), and subcutaneous or epidural infusion of medications directly into an intraoperative site or s Recall
Class II: A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Class II recall indicates potential for temporary health effects.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver continuous medication for periods of 30 minutes up to 7 days. They are designed for medication therapies requiring continuous infusion. The infusion pumps can be utilized for slow, continuous delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), and subcutaneous or epidural infusion of medications directly into an intraoperative site or s
Brand
Baxter Healthcare Corp.
Lot Codes / Batch Numbers
*** 1) Product Code: 2C1008KP, Lots: 08M036, 08N041, 09A030, 09B044, 09C053, 09D018, 09E019, 09F005, *** 2) Product Code: 2C1009KP, Lots: 08M016, 08M035, 08M066, 09A010, 09A052, 09A065, 09A080, 09B038, 09B048, 09B060, 09B074, 09C03, 09C016, 09C044, 09C060, 09D011, 09D042, 09D049, 09D054, 09E021, 09E043, 09F002, *** 3) Product Code: 2C1063KP, Lots: 08K078, 08M014, 08M060, 08N022, 08N053, 08N074, 09A032, 09A057, 09B006, 09B014, 09C007, 09C017, 09C054, 09C077, 09D015, 09D041, 09E012, 09E018, 09E022, 09E044, *** 4) Product Code: 2C1087KP, Lots: 08K074, 08M050, 08N051, 09A053, 09B063, 09C041, 09D044, 09E073, *** 5) Product Code: 2C1156KP, Lots: 08M063, 09C042
Products Sold
*** 1) Product Code: 2C1008KP; Lots: 08M036, 08N041, 09A030, 09B044, 09C053, 09D018, 09E019, 09F005; *** 2) Product Code: 2C1009KP; Lots: 08M016, 08M035, 08M066, 09A010, 09A052, 09A065, 09A080, 09B038, 09B048, 09B060, 09B074, 09C03, 09C016, 09C044, 09C060, 09D011, 09D042, 09D049, 09D054, 09E021, 09E043, 09F002; *** 3) Product Code: 2C1063KP; Lots: 08K078, 08M014, 08M060, 08N022, 08N053, 08N074, 09A032, 09A057, 09B006, 09B014, 09C007, 09C017, 09C054, 09C077, 09D015, 09D041, 09E012,09E018, 09E022, 09E044; *** 4) Product Code: 2C1087KP; Lots: 08K074, 08M050, 08N051, 09A053, 09B063, 09C041, 09D044, 09E073; *** 5) Product Code: 2C1156KP; Lots: 08M063, 09C042
Baxter Healthcare Corp. is recalling Baxter's elastomeric infusion pumps are non-electronic, ambulatory, disposable devices that deliver due to There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investiga. This is a Class I recall, indicating a high risk of serious health consequences or death. Based on FDA medical device enforcement report.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There are reported incidents of leakage on the INFUSOR and FOLFUSOR pumps at the connection of the blue winged cap and the distal male Luer. Investigation of these reports indicates the incidence of leakage has only occurred after pumps are filled, primed, re-capped (using the blue winged Luer cap) and stored for a period of time, typically overnight. In addition, the reported instances of leakage
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026