Baxter Healthcare Corp. Rt. Baxter 6060 Multi-Therapy Infusion Pump, product code 2M9832, 606000-40, 606000-40L, 606000-40I; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Baxter 6060 Multi-Therapy Infusion Pump, product code 2M9832, 606000-40, 606000-40L, 606000-40I; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore
Brand
Baxter Healthcare Corp. Rt.
Lot Codes / Batch Numbers
all serial numbers
Products Sold
all serial numbers
Baxter Healthcare Corp. Rt. is recalling Baxter 6060 Multi-Therapy Infusion Pump, product code 2M9832, 606000-40, 606000-40L, 606000-40I; Man due to There is an overinfusion condition in the Intermittent mode where the pump may continue to deliver at the higher dose rate instead of switching to the. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is an overinfusion condition in the Intermittent mode where the pump may continue to deliver at the higher dose rate instead of switching to the lower keep open rate once the first dose is completed.
Recommended Action
Per FDA guidance
Baxter sent the 6/13/05 Urgent Device Correction letter to all Baxter 6060 and 6060E Infusion Pump customers to inform them of a software anomaly which allows the pump to overinfuse in the Intermittent mode of the 6060 pump if the following specific sequence of events occur: * the infusion is running in intermittent mode; * the pump is delivering at the Keep Open (KO) rate; * the pump is then powered off and on; * the clinician initiates new programming of the pump by selecting NO at the 'Resume KO Dose 1?' prompt and selecting YES at the 'Yes to Program' prompt; * the clinician powers the pump off and on again; * and the clinician selects YES at the 'Resume KO Dose 1?' prompt. Until the software upgrade is available, Baxter recommended that the accounts notify healthcare providers of the potential for an overinfusion condition in Intermittent mode if the above sequence is performed during an infusion, and to utilize the Lockout mode after programming the pump. Any questions were directed to Baxter''s Medication Delivery Services at 1-800-843-7867. The letter also informed the accounts of two patient deaths due to medication delivery errors resulting from improper programming of the infusion parameters into the 6060 and 6060E pumps, and reports where the infusion parameters were modified by untrained individuals, which may result in overinfusion or underinfusion conditions. The letter stressed that programming of the pump must be performed only by trained clinicians, or healthcare providers trained and deemed competent by trained clinicians or under the direct supervision of trained clinicians; the Lockout mode feature of the pump should be used to prevent unauthorized changes to the pump programming; and the pump security codes, which disable the Lockout mode, must not be released to anyone other than trained healthcare professionals. A copy of the warning from the Operator''s Manual was included with the letter. Any questions were referred to the Center fo
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026