Baxter Healthcare Corp. Rt. Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A.
Brand
Baxter Healthcare Corp. Rt.
Lot Codes / Batch Numbers
serial numbers 15060001AB, 15060002AB, 15060004AB/15060013AB, 15060015AB/15060021AB, 15060023AB/15060073AB, 15060075AB, 15060077AB/15060116AB, 15060116AB, 15060118AB/15060126AB, 15060128AB/15060130AB, 15060132AB/15060172AB, 15060174AB/15060209AB, 15070003AB/15070006AB, 15070008AB/15070018AB, 15070021AB, 15070023AB, 15070026AB/15070034AB, 15070037AB15070041AB, 15070044AB/15070047AB, 15070049AB/15070052AB, 15070056AB/15070059AB, 15070066AB, 15070069AB, 15070071AB, 15070072AB, 15070076AB, 15070077AB, 15070079AB, 15070081AB, 15070082AB, 15070084AB, 15070086AB, 15070087AB, 15070092AB, 15070094AB/15070096AB, 15070099AB, 15070101AB, 15070109AB, 15070111AB/15070115AB, 15070117AB, 15070119AB, 15070121AB, 15070122AB, 15070124AB, 15070126AB/15070128AB, 15070130AB/15070132AB, 15070138AB/15070140AB, 15070144AB
Products Sold
serial numbers 15060001AB, 15060002AB, 15060004AB/15060013AB, 15060015AB/15060021AB, 15060023AB/15060073AB, 15060075AB, 15060077AB/15060116AB, 15060116AB, 15060118AB/15060126AB, 15060128AB/15060130AB, 15060132AB/15060172AB, 15060174AB/15060209AB, 15070003AB/15070006AB, 15070008AB/15070018AB, 15070021AB, 15070023AB, 15070026AB/15070034AB, 15070037AB15070041AB, 15070044AB/15070047AB, 15070049AB/15070052AB, 15070056AB/15070059AB, 15070066AB, 15070069AB, 15070071AB, 15070072AB, 15070076AB, 15070077AB, 15070079AB, 15070081AB, 15070082AB, 15070084AB, 15070086AB, 15070087AB, 15070092AB, 15070094AB/15070096AB, 15070099AB, 15070101AB, 15070109AB, 15070111AB/15070115AB, 15070117AB, 15070119AB, 15070121AB, 15070122AB, 15070124AB, 15070126AB/15070128AB, 15070130AB/15070132AB, 15070138AB/15070140AB, 15070144AB
Baxter Healthcare Corp. Rt. is recalling Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medicati due to A non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which c. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which could increase the potential for the pump to fail during use and alarm.
Recommended Action
Per FDA guidance
Baxter sent Urgent Device Correction letters dated 2/7/06 to the direct accounts on the same date via first class mail. The letters informed the accounts of the potential for the affected pumps to be susceptible to electro-static discharge, which could increase the potential to fail during use, causing the pump to stop infusing and issue an audible and visual alarm. The accounts were asked to immediately take the pump out of service, contact Baxter at 1-800-843-7867 to arrange for the return of the pumps in order to replace the non-conforming gasket, and complete and return the enclosed reply form confirming receipt of the letter. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.