Baxter Healthcare Corp. Rt. Baxter Colleague Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore. Product Codes: 2M8151, 2M8161, 2M9161, DNM8151, DNM9161, BRM8151, BRM8151T, BRM8151RT, PNM8151, PNM8151T, PNM8151RT, CNM8151, FKM8151, GNM8151, HNM8151, TRM8151 and WNM8151. Electronic infusion pumps are indicated for continuous or intermittent delivery of solution through clinically acceptable routes of administration, such as Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Baxter Colleague Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore. Product Codes: 2M8151, 2M8161, 2M9161, DNM8151, DNM9161, BRM8151, BRM8151T, BRM8151RT, PNM8151, PNM8151T, PNM8151RT, CNM8151, FKM8151, GNM8151, HNM8151, TRM8151 and WNM8151. Electronic infusion pumps are indicated for continuous or intermittent delivery of solution through clinically acceptable routes of administration, such as
Brand
Baxter Healthcare Corp. Rt.
Lot Codes / Batch Numbers
All product numbers and all serial numbers.
Products Sold
All product numbers and all serial numbers.
Baxter Healthcare Corp. Rt. is recalling Baxter Colleague Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medicatio due to Baxter has identified failure codes that lead to an interruption of therapy, hazards associated with improper cleaning, and damaged battery messages f. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Baxter has identified failure codes that lead to an interruption of therapy, hazards associated with improper cleaning, and damaged battery messages for the Colleague Infusion Pumps.
Recommended Action
Per FDA guidance
Baxter Healthcare Corporation sent an "Urgent Device Correction" letter dated January 23, 2009 to all Colleague Infusion Pump customers, to the attention of the Vice President of Nursing with copies to the Chief Operating Officer, Director of Biomedical Engineering and Director of Risk Management to advise them of failure codes that lead to an interruption of therapy, hazards associated with improper cleaning, and damaged battery messages associated with the Colleague pumps. The letters listed the failure codes involved that lead to an interruption of therapy which could lead to serious injury and/or death; advised institutions to have contingency plans to verify that back-up pumps are available; and provided steps for addressing an interruption of therapy with any failure code. The letters also provided information about the hazards associated with improper cleaning of the pumps and instructions for addressing damaged battery messages. Customers were instructed to complete the attached Customer Reply Form acknowledging receipt of the letter and fax it to 888-569-1585. Direct questions to your local Baxter Representative or to Baxter Global Technical Services by calling 1-800-843-7867, Prompt 2, between 7AM and 5 PM Central Standard Time (CST).