Baxter Healthcare Corp. Rt. Baxter Colleague Triple Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore; product codes 2M8153, 2M8153R, BRM8153(Brazilian Portugese), DNM8153(French), HNM8153(German), PNM8153(Spanish), CNM8153(Swedish), GNM8153(Danish), TRM8153(Turkish), WNM8153(Dutch) ,2M8163, 2M8163R Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Baxter Colleague Triple Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore; product codes 2M8153, 2M8153R, BRM8153(Brazilian Portugese), DNM8153(French), HNM8153(German), PNM8153(Spanish), CNM8153(Swedish), GNM8153(Danish), TRM8153(Turkish), WNM8153(Dutch) ,2M8163, 2M8163R
Brand
Baxter Healthcare Corp. Rt.
Lot Codes / Batch Numbers
all serial numbers
Products Sold
all serial numbers
Baxter Healthcare Corp. Rt. is recalling Baxter Colleague Triple Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication due to Interruption of therapy due to battery undercharging, the generation of air-in-line alarms due to IV administration set tugging, and gearbox wear; and. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Interruption of therapy due to battery undercharging, the generation of air-in-line alarms due to IV administration set tugging, and gearbox wear; and underinfusion due to misalignment of the pump head components.
Recommended Action
Per FDA guidance
Baxter sent the 12/13/05 Urgent Device Correction letter to all Colleague Infusion Pump customers, to the attention of the Director of Nursing and the Biomedical Engineer, via first class mail to alert them to additional problems with the pump that may interrupt therapy. These issues include battery undercharging, generation of false air detected alarms due to IV administration set tugging, gearbox wear, underinfusion and non-detection of upstream occlusions. The letters included mitigating instructions to reduce the occurrence of these potential problems, and included a copy of the Battery Usage Guide, Infusion Management Guide, and the March 17, 2005 Buretrol Set Urgent Device Correction letter to be provided to all users of the Colleague pump. Baxter informed the accounts that they would be notified when the new release of the Colleague Volumetric Infusion Pump Operator''s Manual is available. Any questions were directed to Baxter''s Medication Delivery Services at 1-800-843-7867.