Baxter Healthcare Corp. Rt. Colleague CX Single Channel Volumetric Infusion Pumps, product code 2M8161and 2M8161R; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in Singapore; color display screen Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Colleague CX Single Channel Volumetric Infusion Pumps, product code 2M8161and 2M8161R; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in Singapore; color display screen
Brand
Baxter Healthcare Corp. Rt.
Lot Codes / Batch Numbers
all serial numbers
Products Sold
all serial numbers
Baxter Healthcare Corp. Rt. is recalling Colleague CX Single Channel Volumetric Infusion Pumps, product code 2M8161and 2M8161R; Baxter Health due to A software anomaly causes a failure alarm code 12:303 which audibly alarms and stops the function of all channels in use. This causes an interruption. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A software anomaly causes a failure alarm code 12:303 which audibly alarms and stops the function of all channels in use. This causes an interruption in therapy, which may result in a risk to the patient.
Recommended Action
Per FDA guidance
Baxter sent Urgent Device Correction letters dated 11/4/04 to the direct accounts to the attention of the Director of Biomedical Engineering on the same date. The letters informed the accounts of the 12:303 failure and possible interruption of patient therapy when it occurs. They were informed that new software was available to address the failure alarm code with the infusion pumps, and that the software also included updates to the predefined label library and the addition of a text message 'Close regulating roller clamp' to the manual tube release pop-up message. The software is being installed on all single and triple channel Colleague infusion pumps as they are processed through Baxter''s depot repair centers for other repair actions or routine maintenance. An Operator''s Manual Addendum was included with the letter to reflect the changes to the label library and the manual tube release pop-up message. The accounts were requested to complete the attached reply form and return it via fax to Baxter. The completed form acknowledges receipt of the recall letter, verifies the model numbers and serial numbers of the pumps at the facility and indicates the number of revised software installation kits needed by self-service customers. Any questions were directed to Baxter at 1-800-843-7867.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.