Baxter Healthcare Corp. Rt. Configuration Transfer Cable; an accessory for the Colleague Infusion pump, Product code 2M8155; Baxter Healthcare Corporation, Deerfield, IL Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Configuration Transfer Cable; an accessory for the Colleague Infusion pump, Product code 2M8155; Baxter Healthcare Corporation, Deerfield, IL
Brand
Baxter Healthcare Corp. Rt.
Lot Codes / Batch Numbers
Product code 2M8155, batch #50604
Products Sold
Product code 2M8155, batch #50604
Baxter Healthcare Corp. Rt. is recalling Configuration Transfer Cable; an accessory for the Colleague Infusion pump, Product code 2M8155; Bax due to The Configuration Tranfer Cable for the Colleague Infusion Pumps has incorrect thumb screws on it., not allowing firm connections of the cable to the . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Configuration Tranfer Cable for the Colleague Infusion Pumps has incorrect thumb screws on it., not allowing firm connections of the cable to the pump to transfer data to the biomedical engineer's diagnostic device.
Recommended Action
Per FDA guidance
Recall letters dated 3/2/05 were sent to all customers receiving Configuration Transfer Cables between June and November 2004. The letters were sent to the attention of the Biomedical Engineer, and informed them that the thumbscrews were not manufactured according to Baxter''s specifications and were too narrow to secure the cable to the pump. The accounts were requested to return their Configuration Transfer Cables for replacement.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026