Baxter Healthcare Corporation ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abacus V3.1 SE), 8300-0169 (Abacus V3.1 ME), 8300-0191 (Abacus V3.2 CE), 8300-0192 (Abacus V3.2 SE), 8300-0193 (Abacus V3.2 ME), 8300-3391 (Abacus V3.3 CE), 8300-3392 (Abacus V3.3 SE), and 8300-3393 (Abacus V3.3 ME). Commonly used with ExactaMix automated compounding devices (ExactaMix 1200 and ExactaMix 2400). Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abacus V3.1 SE), 8300-0169 (Abacus V3.1 ME), 8300-0191 (Abacus V3.2 CE), 8300-0192 (Abacus V3.2 SE), 8300-0193 (Abacus V3.2 ME), 8300-3391 (Abacus V3.3 CE), 8300-3392 (Abacus V3.3 SE), and 8300-3393 (Abacus V3.3 ME). Commonly used with ExactaMix automated compounding devices (ExactaMix 1200 and ExactaMix 2400).
Brand
Baxter Healthcare Corporation
Lot Codes / Batch Numbers
Product Code 8300-3391, UDI (01)05413765577345(10)3.3.2.1, Release date 20-Sep-2019, Product Code 8300-3392, UDI (01)05413765577352(10)3.3.2.1, Release date 20-Sep-2019, Product Code 8300-3393, UDI (01)05413765577369(10)3.3.2.1, Release date 20-Sep-2019, Product Codes: 8300-0167, V3.1 CE, Release date 20-May-2013, 8300-0168, V3.1 SE, Release date 20-May-2013, 8300-0169, V3.1 ME, Release date 20-May-2013, 8300-0191, V3.2 CE, Release date 17-Dec-2015, 8300-0192, V3.2 SE, Release date 17-Dec-2015, 8300-0193, V3.2 ME, Release date 17-Dec-2015, No UDIs
Products Sold
Product Code 8300-3391, UDI (01)05413765577345(10)3.3.2.1, Release date 20-Sep-2019; Product Code 8300-3392, UDI (01)05413765577352(10)3.3.2.1, Release date 20-Sep-2019; Product Code 8300-3393, UDI (01)05413765577369(10)3.3.2.1, Release date 20-Sep-2019; Product Codes: 8300-0167, V3.1 CE, Release date 20-May-2013, 8300-0168, V3.1 SE, Release date 20-May-2013, 8300-0169, V3.1 ME, Release date 20-May-2013, 8300-0191, V3.2 CE, Release date 17-Dec-2015, 8300-0192, V3.2 SE, Release date 17-Dec-2015, 8300-0193, V3.2 ME, Release date 17-Dec-2015, No UDIs
Baxter Healthcare Corporation is recalling ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 due to There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.
Recommended Action
Per FDA guidance
Baxter sent an Urgent Medical Device Correction notice on 06/22/2022 by letter. The notice explained the problem and the hazard and recommended the following: customers refrain from making updates to bag label templates and to contact Baxter Technical Support if updates to bag label templates are needed. Please continue to follow the instructions in the Abacus User and Abacus Configuration Guides when creating an order, and have a pharmacist thoroughly review all order outputs including the printed label for accuracy and safety. As mentioned above, please ensure your facility processes include pharmacy checks during the compounding process as well as a nursing check. Baxter will communicate to customers when the software upgrade becomes available. At that time additional instructions to obtain the software upgrade will be provided.