Baxter Healthcare Corporation EXACTAMED Oral Dispensers, 5mL, AMBER: PHARMACY PACK, REF H9388505, x500 PHARMACY PACK, REF H9388105, x100 COST CONTAINMENT, REF H93884205, x1500 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
EXACTAMED Oral Dispensers, 5mL, AMBER: PHARMACY PACK, REF H9388505, x500 PHARMACY PACK, REF H9388105, x100 COST CONTAINMENT, REF H93884205, x1500
Brand
Baxter Healthcare Corporation
Lot Codes / Batch Numbers
REF H9388105, Lot Numbers: B301N017P, B301N070P, B302M338P, B304N045P REF H9388505, Lot Numbers: B301N023P, B301N049P, B302N077P REF H93884205, Lot Numbers: B301N056P, B302M333P, B303N042P, B307N015P
Products Sold
REF H9388105, Lot Numbers: B301N017P, B301N070P, B302M338P, B304N045P REF H9388505, Lot Numbers: B301N023P, B301N049P, B302N077P REF H93884205, Lot Numbers: B301N056P, B302M333P, B303N042P, B307N015P
Baxter Healthcare Corporation is recalling EXACTAMED Oral Dispensers, 5mL, AMBER: PHARMACY PACK, REF H9388505, x500 PHARMACY PACK, REF H9388 due to EXACTAMED 5 mL Oral Dispenser Pharmacy Packs may contain both 5 mL and 10 mL oral dispensers.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
EXACTAMED 5 mL Oral Dispenser Pharmacy Packs may contain both 5 mL and 10 mL oral dispensers.
Recommended Action
Per FDA guidance
The firm disseminated a safety alert to their consignees on 06/27/2019 by letter. The letter explained the 5mL syringes were commingled with 10mL syringes and directed the consignee separate the 5mL syringes from the 10mL syringes.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026