Baxter Healthcare Corporation HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053, P7501A000054, P7501A000055, P7501A001851, P7501A001972, P7501A001973, and P7501A001974 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053, P7501A000054, P7501A000055, P7501A001851, P7501A001972, P7501A001973, and P7501A001974
Brand
Baxter Healthcare Corporation
Lot Codes / Batch Numbers
UDI/DI 00887761966925, all serial numbers
Products Sold
UDI/DI 00887761966925, all serial numbers
Baxter Healthcare Corporation is recalling HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P750 due to Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator thr. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock.
Recommended Action
Per FDA guidance
Baxter issued an Urgent Medical Device Correction notice to its consignees on 10/04/2023 by USPS first class mail. The notice explained the potential problem and hazard. Baxter will be correcting this issue by installing a conductive drag chain on the Progressa+ bed frame to prevent the electrostatic charge buildup. A Baxter representative will contact your facility to schedule a device correction to add the conductive drag chain. Please note you will be receiving this correction from Baxter at no charge. All units purchased in the future will have the drag chain.