Baxter Healthcare Corporation NaviCare Nurse Call/Voalte Nurse Call Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
NaviCare Nurse Call/Voalte Nurse Call
Brand
Baxter Healthcare Corporation
Lot Codes / Batch Numbers
software versions: 3.9.100, 3.9.101, 3.9.102, 3.9.200, 3.9.201, and 3.9.300
Products Sold
software versions: 3.9.100, 3.9.101,3.9.102, 3.9.200, 3.9.201, and 3.9.300
Baxter Healthcare Corporation is recalling NaviCare Nurse Call/Voalte Nurse Call due to An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalt. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.
Recommended Action
Per FDA guidance
The firm disseminated an Urgent Medical Device Correction Notice beginning on 05/31/2022 by letter. The letter described the problem and requested the following actions be taken by the customer: "Identify if your Nurse Call System utilizes one of the impacted wireless integrations identified. If so, please share the information with potential users in your organization. Hillrom recommends reviewing the wireless system settings with your wireless provider to determine if there are configurable options for call acceptance only. Complete the enclosed response form and return to hillromCARY021@sedgwick.com within two weeks of receipt of this notice, to confirm you have received this communication and understand the potential issue." Persons with questions were directed as follows: "If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support, available twenty-four hours a day, seven days a week, at 1-800-445-3720, option 3, or hrccwssupport@hillrom.com"
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.