Novum IQ Large Volume Pump (Baxter) – device correction (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Novum IQ LVP (Large Volume Pump), Product Code 40700BAXUS, infusion pump
Brand
Baxter Healthcare Corporation
Lot Codes / Batch Numbers
UDI/DI 05413765851797, All Serial Numbers
Products Sold
UDI/DI 05413765851797, All Serial Numbers
Baxter Healthcare Corporation is recalling Novum IQ LVP (Large Volume Pump), Product Code 40700BAXUS, infusion pump due to Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) d. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).
Recommended Action
Per FDA guidance
Baxter issued an "URGENT MEDICAL DEVICE CORRECTION" to its consignees on 08/04/2025 via USPS first class mail. The notice described the problem, hazard involved, and actions to be Taken by customers: 1. Users should rely on their clinical judgment. When providing patient care involving high risk medications and/or critical illness, users should consider utilizing an alternate pump, if available and appropriate for the patient's treatment. 2. If an alternate pump is not available or appropriate for patient care you may continue to use the Novum IQ LVP and Novum IQ Syringe Pump consistent with the following actions: - If a blank Run screen is displayed during an infusion on the Novum IQ LVP, the user should open the door, press the tubing into load point 3, then load point 4, to turn the light green, and close the door. If a blank Run screen is displayed during an infusion on the Novum IQ Syringe Pump, the user should unload, then reload the syringe. If these actions do not resolve the blank screen, both LVP and Syringe Pump users should power-cycle the pump, select "Yes" to the new patient prompt, and reprogram the infusion. - To prevent a false motor movement system error, do not program the Novum IQ syringe pump to infuse below the Minimum Recommended Flow Rate for a syringe size of 10 mL or greater. 3. Acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. 4. Please forward a copy of this communication to any departments within your institution who use the affected product. 5. Print and post Attachment A in areas frequented by potential users. For general questions regarding this communication, or if you experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867 (select option 2, then option 2 again) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA