Spectrum IQ Infusion Pump (Baxter) – Grease Breakdown (2025)
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Spectrum IQ Infusion Pump, Product Code 3570009
Brand
Baxter Healthcare Corporation
Lot Codes / Batch Numbers
UDI-DI: 00085412610900, Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271
Products Sold
UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271
Baxter Healthcare Corporation is recalling Spectrum IQ Infusion Pump, Product Code 3570009 due to Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickl. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
Recommended Action
Per FDA guidance
Customers were contacted via telephone beginning 7/14/25. The included information about the product, the issue and the action to be taken by the customers. Baxter Healthcare is requesting the return of the impacted product to perform all tests required and confirm that the product perform as expected. If you have any questions, phone (Toll Free): 800-422-9837 or 847-948-4770; hours of operation: Monday through Friday from 8:00 am to 5:00 pm CST
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
AZ, FL
Page updated: Jan 10, 2026