Baxter Healthcare Corporation Voalte Nurse Call System provides a comprehensive communication and information system that places patient calls, staff calls, and emergency calls. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Voalte Nurse Call System provides a comprehensive communication and information system that places patient calls, staff calls, and emergency calls.
Brand
Baxter Healthcare Corporation
Lot Codes / Batch Numbers
Internal Part Number: 4.0., UDI/GTIN: 00887761995079, P2519NNC3A24 model number of impacted RubyTech Switch
Products Sold
Internal Part Number: 4.0.; UDI/GTIN: 00887761995079; P2519NNC3A24 model number of impacted RubyTech Switch
Baxter Healthcare Corporation is recalling Voalte Nurse Call System provides a comprehensive communication and information system that places due to Firm discovered a firmware memory leak with a supplier-manufactured component.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Firm discovered a firmware memory leak with a supplier-manufactured component.
Recommended Action
Per FDA guidance
The firm, Hillrom, issued an "Urgent Medical Device Correction Notice", dated 04/13/2022, to impacted consignees via email. The consignee is instructed to identify if they have the impacted product component, Part P2519NNC3A2 and also distribute the notification to personnel as well ensure that the users monitor for system alerts. Notify Hillrom Technical support if there is a loss of connectivity to the nurse call server. Note, error indications will display on multiple user interfaces. Central nurses' stations will state the room(s) is down (disconnected). In patient room(s), graphical audio stations will display clear message stating Station down. Contact Nurse directly . In addition, standard audio stations will have an LED Error state indicator. Complete and return the response form to william.knouse@hillrom.com. Hillrom is testing a firmware update for the Power over Ethernet (PoE) switch and anticipates availability in April 2022. Upon receipt of the response form, Hillrom will contact you to schedule an update to your system once firmware is available. If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support at +1 919 854 3361 or william.knouse@hillrom.com