Baxter Healthcare Renal Div Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.
Brand
Baxter Healthcare Renal Div
Lot Codes / Batch Numbers
all serial numbers
Products Sold
all serial numbers
Baxter Healthcare Renal Div is recalling Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, Re due to The Meridian pump is less likely to detect small pressure changes, which may indicate the presence of a post-pump tubing kink, when using pre-pump art. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Meridian pump is less likely to detect small pressure changes, which may indicate the presence of a post-pump tubing kink, when using pre-pump arterial pressure monitoring.
Recommended Action
Per FDA guidance
Urgent Device Correction letters dated 9/6/06 were sent to the Meridian accounts on the same date via UPS second day, to the attention of the Hemodialysis Administrator. The letters informed the accounts of the potential risk associated with a post-pump bloodline kink when using pre-pump arterial pressure monitoring. The accounts were provided with a revised Meridian Operator''s Manual which includes the following warnings: 'The lower venous limit should be set as close to the currently displayed venous pressure as conditions allow.' and 'Reliance on the pressure alarms should not substitute for clinical monitoring of the patient for adverse events.' Any technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 3, option 2, choice 2. General questions were directed to The Center for One Baxter at 1-800-422-9837.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026