PE Centrifuge Temperature Controlled, REF: B36366, a comp... (Beckman Coulter Biomedical GmbH Sauerbruchstr. 50 Munchen Germany) – metal contamination (2024)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
Brand
Beckman Coulter Biomedical GmbH Sauerbruchstr. 50 Munchen Germany
Lot Codes / Batch Numbers
UDI: (01)15099590673567/ All serial numbers
Products Sold
UDI: (01)15099590673567/ All serial numbers
Beckman Coulter Biomedical GmbH Sauerbruchstr. 50 Munchen Germany is recalling PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express. due to The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and en. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.
Recommended Action
Per FDA guidance
On July 18, 2024, Beckman Coulter issued a "Urgent Medical Device Recall" notification to affected consignees. Beckman Coulter asked consignees to take the following actions: 1. follow your laboratory procedure to handle biohazard materials and contact your Beckman Coulter representative for servicing your Power Express system. 2. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. 3. respond within 10 days in one of the following ways: . Electronically, if you received this communication via email. . Manually, complete and return the enclosed Response Form. 4. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026