Beckman Coulter Inc Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System, for in vitro diagnostic use. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System, for in vitro diagnostic use.
Brand
Beckman Coulter Inc
Lot Codes / Batch Numbers
Part Number A48196/6845C, Instrument part numbers 81600 and 81600N
Products Sold
Part Number A48196/6845C; Instrument part numbers 81600 and 81600N
Beckman Coulter Inc is recalling Access Power Supply Assembly Sled Part Number: A48196/6845C; Component of: Access and Access 2 Immun due to Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. T. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Electrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.
Recommended Action
Per FDA guidance
Beckman Coulter mailed a Product Corrective Action (PCA) letter on October 17, 2007 to all Access and Access 2 lmmunoassay Systems, SYNCHRON LXi 725 System, and UniCel DxC 600i System customers. The customers were informed that during manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering. The letter referenced a proactive initiative to inspect the power supply sled assemblies on all instruments referenced. A response form was included.