Beckman Coulter Inc Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of insulin levels in human serum and plasma (EDTA) using the Access Immunoassay Systems. Beckman Coulter, Inc., Chaska, MN 55318. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of insulin levels in human serum and plasma (EDTA) using the Access Immunoassay Systems. Beckman Coulter, Inc., Chaska, MN 55318.
Brand
Beckman Coulter Inc
Lot Codes / Batch Numbers
Part Number 33410, Lot Numbers: 714362 (Exp dt. 3/31/09), 716129 (Exp dt. 6/30/09), 718258 (Exp dt. 7/31/09), 519096 (Exp dt. 10/31/07), 521487 (Exp dt. 2/28/08), 616274 (Exp dt. 7/31/08), 617890 (Exp dt. 7/31/08), 617891 (Exp dt. 7/31/08) and 620204 (Exp dt. 12/31/08)
Products Sold
Part Number 33410, Lot Numbers: 714362 (Exp dt. 3/31/09), 716129 (Exp dt. 6/30/09), 718258 (Exp dt. 7/31/09), 519096 (Exp dt. 10/31/07), 521487 (Exp dt. 2/28/08), 616274 (Exp dt. 7/31/08), 617890 (Exp dt. 7/31/08), 617891 (Exp dt. 7/31/08) and 620204 (Exp dt. 12/31/08)
Beckman Coulter Inc is recalling Access Ultrasensitive Insulin Assay Kit, Part Number 33410. The product is a paramagnetic particle, due to False negative results when used to test serum samples (as compared to plasma samples).. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
False negative results when used to test serum samples (as compared to plasma samples).
Recommended Action
Per FDA guidance
On September 19, 2007, Beckman Coulter mailed to its consignees an Urgent Product Corrective Action Letter informing them that the affected lots of the product were giving false negative results when used to test serum samples (as compared to plasma samples). The letter instructed consignees to discontinue using the affected lots of the product, and to re-test using an EDTA plasma sample if there was any discordance between the results of the serum sample (if it was obtained using an affected lot of the product) as compared with the clinical presentaion of the patient and the results of other diagnostic tests. The letter also instructed consignees to return, within 10 days, the Response Sheet .