UniCel DxI 800 Analyzer (Beckman) – motion error risk (2025)
This recall may present a risk under certain conditions.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer
Brand
Beckman Coulter, Inc.
Lot Codes / Batch Numbers
973100, UDI/DI 15099590265373 (Chaska), Serial numbers: 610458, 610455, 610679, 610579, 610432, 610430, 610434, 610664, 610682, 610765, 610500, 610541, 610544, 610545, 610562, 610547, 610585, 610595, 610475, 610453, 610781, 610580, 610444, 610704, 610586, 610684, 610589, 610588, 610493, 610659, 610570, 610498, 610748, 610496, 610560, 610573, 610574, 610577, 610601, 610651, 610680, 610642, 610641, 610683, 610747, 610468, 610772, 610441, 610476, 610489, 610488, 610492, 610532, 610534, 610533, 610535, 610640, 610457, 610464, 610790, 610559, 610687, 610571, 610552, 610555, 610582, 610531, 610568, 610594, 610773, 610490, 610668, 610784, 610812, 610556, 610550, 610600, 610565, 610558, 610557, 610551, 610527, 610685, 610730, 610686, 610494, 610753, 610590, 610593, 610546, 610724, 610538, 610540, 610439, 610491, 610581, 610649, 610466, 610789, 610723, 610575, 610443, 610776, 610734, 610518, 610519, 610517, 610520, 610563, 610569, 610787, 610459, 610775, 610572, 610561, 610597, 610445, 610440, 610567, 610596, 610497, 610774, 610454, 610426, 610431, 610442, 610433, 610435, 610539, 610578, 610509, 610418, 610483, 610484, 610515, 610425, 610503, 610510, 610452, 610526, 610602, 610481, 610487, 610482, 610507, 610505, 610506, 610514, 610499, 610504, 610511, 610423, 610524, 610525, 610512, 610564, 610424, 610513, 610480, 610516, 610521, 610419, 610415, 610416, 610456, 610461, 610460, 610472, 610470, 610477, 610471, 610463, 610462, 610467, 610469, 610465, 610587, 610639, 610614, 610616, 610608, 610605, 610645, 610634, 610623, 610626, 610615, 610633, 610622, 610609, 610621, 610672, 610678, 610674, 610671, 610667, 610673, 610675, 610677, 610650, 610662, 610692, 610646, 610693, 610652, 610661, 610663, 610654, 610655, 610689, 610653, 610656, 610657, 610694, 610688, 610658, 610660, 610725, 610729, 610732, 610731, 610719, 610726, 610703, 610709, 610718, 610715, 610717, 610716, 610698, 610702, 610714, 610711, 610705, 610706, 610697, 610710, 610755, 610757, 610754, 610756, 610767, 610766, 610769, 610768, 610785, 610788, 610800, 610793, 610794, 610795, 610801, 610814, 610811, 610758, 610786, 610744, 610804, 610736, 610737, 610742, 610750, 610802, 610806, 610791, 610759, 610762, 610738, 610739, 610740, 610741, 610792, 610805, 610808, 610733, 610761, 610763, 610798, 610803, 610735, 610810, 610760, 610749, 610799, 610809, 610764, 610745, 610743, 610746, 610829, 610832, 610833, 610830, 610831, 610827, 610828, 610825, 610826, 610822, 610807, 610813, 610821. UDI/DI 16954701000412 (Suzhou), Serial Numbers: 630104, 630105, 630160, 630159, 630162, 630110, 630140, 630175, 630143, 630103, 630108, 630174, 630099, 630112, 630113, 630136, 630107, 630102, 630100, 630106, 630115, 630114, 630132, 630111, 630101, 630109, 630146, 630139, 630116, 630126, 630117, 630118, 630120, 630119, 630121, 630122, 630123, 630124, 630125, 630129, 630128, 630130, 630131, 630134, 630135, 630137, 630141, 630142, 630133, 630127, 630164, 630172, 630173, 630144, 630156, 630145, 630157, 630158, 630155, 630154, 630149, 630152, 630170, 630171, 630150, 630151, 630153, 630163, 630148, 630147, 630138, 630194, 630197, 630202, 630203, 630178, 630184, 630166, 630186, 630182, 630176, 630205, 630179, 630180, 630181, 630195, 630187, 630161, 630169, 630167, 630165, 630193, 630191, 630192, 630190, 630177, 630188, 630183, 630196, 630199, 630168, 630219, 630208, 630217, 630230, 630223, 630214, 630210, 630235, 630232, 630201, 630226, 630227, 630218, 630233, 630221, 630222, 630216, 630215, 630206, 630207, 630211, 630231, 630209, 630224, 630234, 630228, 630220, 630225, 630229, 630204, 630212, 630213, 630259, 630260, 630252, 630281, 630282, 630272, 630265, 630279, 630256, 630255, 630269, 630242, 630248, 630241, 630264, 630275, 630270, 630249, 630253, 630271, 630267, 630278, 630254, 630263, 630243, 630246, 630273, 630280, 630238, 630251, 630257, 630258, 630274, 630240, 630244, 630247, 630262, 630237, 630266, 630236, 630261, 630245, 630307, 630300, 630301, 630315, 630284, 630319, 630327, 630296, 630292, 630304, 630305, 630311, 630310, 630312, 630324, 630291, 630321, 630323, 630325, 630289, 630316, 630317, 630308, 630306, 630318, 630299, 630314, 630302, 630313, 630290, 630285, 630298, 630293, 630295, 630303, 630320, 630309, 630294, 630287, 630286, 630331, 630288, 630322, 630328, 630337, 630351, 630341, 630345, 630344, 630343, 630342, 630340, 630334, 630330, 630332, 630336, 630338, 630333, 630339, 630347, 630346, 630348, 630326, 630198, 630189, 630276, 630353, 630185, 630352, 630329, 630277, 630283, 630297, 630349, 630200, 630239.
Products Sold
973100, UDI/DI 15099590265373 (Chaska), Serial numbers: 610458, 610455, 610679, 610579, 610432, 610430, 610434, 610664, 610682, 610765, 610500, 610541, 610544, 610545, 610562, 610547, 610585, 610595, 610475, 610453, 610781, 610580, 610444, 610704, 610586, 610684, 610589, 610588, 610493, 610659, 610570, 610498, 610748, 610496, 610560, 610573, 610574, 610577, 610601, 610651, 610680, 610642, 610641, 610683, 610747, 610468, 610772, 610441, 610476, 610489, 610488, 610492, 610532, 610534, 610533, 610535, 610640, 610457, 610464, 610790, 610559, 610687, 610571, 610552, 610555, 610582, 610531, 610568, 610594, 610773, 610490, 610668, 610784, 610812, 610556, 610550, 610600, 610565, 610558, 610557, 610551, 610527, 610685, 610730, 610686, 610494, 610753, 610590, 610593, 610546, 610724, 610538, 610540, 610439, 610491, 610581, 610649, 610466, 610789, 610723, 610575, 610443, 610776, 610734, 610518, 610519, 610517, 610520, 610563, 610569, 610787, 610459, 610775, 610572, 610561, 610597, 610445, 610440, 610567, 610596, 610497, 610774, 610454, 610426, 610431, 610442, 610433, 610435, 610539, 610578, 610509, 610418, 610483, 610484, 610515, 610425, 610503, 610510, 610452, 610526, 610602, 610481, 610487, 610482, 610507, 610505, 610506, 610514, 610499, 610504, 610511, 610423, 610524, 610525, 610512, 610564, 610424, 610513, 610480, 610516, 610521, 610419, 610415, 610416, 610456, 610461, 610460, 610472, 610470, 610477, 610471, 610463, 610462, 610467, 610469, 610465, 610587, 610639, 610614, 610616, 610608, 610605, 610645, 610634, 610623, 610626, 610615, 610633, 610622, 610609, 610621, 610672, 610678, 610674, 610671, 610667, 610673, 610675, 610677, 610650, 610662, 610692, 610646, 610693, 610652, 610661, 610663, 610654, 610655, 610689, 610653, 610656, 610657, 610694, 610688, 610658, 610660, 610725, 610729, 610732, 610731, 610719, 610726, 610703, 610709, 610718, 610715, 610717, 610716, 610698, 610702, 610714, 610711, 610705, 610706, 610697, 610710, 610755, 610757, 610754, 610756, 610767, 610766, 610769, 610768, 610785, 610788, 610800, 610793, 610794, 610795, 610801, 610814, 610811, 610758, 610786, 610744, 610804, 610736, 610737, 610742, 610750, 610802, 610806, 610791, 610759, 610762, 610738, 610739, 610740, 610741, 610792, 610805, 610808, 610733, 610761, 610763, 610798, 610803, 610735, 610810, 610760, 610749, 610799, 610809, 610764, 610745, 610743, 610746, 610829, 610832, 610833, 610830, 610831, 610827, 610828, 610825, 610826, 610822, 610807, 610813, 610821. UDI/DI 16954701000412 (Suzhou), Serial Numbers: 630104, 630105, 630160, 630159, 630162, 630110, 630140, 630175, 630143, 630103, 630108, 630174, 630099, 630112, 630113, 630136, 630107, 630102, 630100, 630106, 630115, 630114, 630132, 630111, 630101, 630109, 630146, 630139, 630116, 630126, 630117, 630118, 630120, 630119, 630121, 630122, 630123, 630124, 630125, 630129, 630128, 630130, 630131, 630134, 630135, 630137, 630141, 630142, 630133, 630127, 630164, 630172, 630173, 630144, 630156, 630145, 630157, 630158, 630155, 630154, 630149, 630152, 630170, 630171, 630150, 630151, 630153, 630163, 630148, 630147, 630138, 630194, 630197, 630202, 630203, 630178, 630184, 630166, 630186, 630182, 630176, 630205, 630179, 630180, 630181, 630195, 630187, 630161, 630169, 630167, 630165, 630193, 630191, 630192, 630190, 630177, 630188, 630183, 630196, 630199, 630168, 630219, 630208, 630217, 630230, 630223, 630214, 630210, 630235, 630232, 630201, 630226, 630227, 630218, 630233, 630221, 630222, 630216, 630215, 630206, 630207, 630211, 630231, 630209, 630224, 630234, 630228, 630220, 630225, 630229, 630204, 630212, 630213, 630259, 630260, 630252, 630281, 630282, 630272, 630265, 630279, 630256, 630255, 630269, 630242, 630248, 630241, 630264, 630275, 630270, 630249, 630253, 630271, 630267, 630278, 630254, 630263, 630243, 630246, 630273, 630280, 630238, 630251, 630257, 630258, 630274, 630240, 630244, 630247, 630262, 630237, 630266, 630236, 630261, 630245, 630307, 630300, 630301, 630315, 630284, 630319, 630327, 630296, 630292, 630304, 630305, 630311, 630310, 630312, 630324, 630291, 630321, 630323, 630325, 630289, 630316, 630317, 630308, 630306, 630318, 630299, 630314, 630302, 630313, 630290, 630285, 630298, 630293, 630295, 630303, 630320, 630309, 630294, 630287, 630286, 630331, 630288, 630322, 630328, 630337, 630351, 630341, 630345, 630344, 630343, 630342, 630340, 630334, 630330, 630332, 630336, 630338, 630333, 630339, 630347, 630346, 630348, 630326, 630198, 630189, 630276, 630353, 630185, 630352, 630329, 630277, 630283, 630297, 630349, 630200, 630239.
Beckman Coulter, Inc. is recalling Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer due to An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
Recommended Action
Per FDA guidance
Beckman Coulter issue an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/01/2025 via USPS and email. The notice explained the issue, potential impact, and requested the following: If Wash PnP and/or Incubator PnP Z-motion errors appear in the analyzer event log, take the following steps: 1. Attempt to reinitialize the analyzer: a. Go to the Diagnostics screen on the User Interface. b. Select Initialize System. 2. If reinitialization is unsuccessful, contact Beckman Coulter Technical Support in your region. Beckman Coulter recommends posting this letter on or near the affected analyzers. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. A field service representative will contact your laboratory to schedule a time to replace the affected parts. If you have any questions regarding this notice, please contact our Customer Support Center website: http://www.beckmancoulter.com