Beckman Coulter Inc Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733), for in vitro diagnostic use. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733), for in vitro diagnostic use.
Brand
Beckman Coulter Inc
Lot Codes / Batch Numbers
Part Number 973100, Serial Numbers: 600469, 601097, 601098, 601144, 601182, 601160, 601188, 601071, 601117, 601114, 601137, 601059, 601039, 601099, 601088, 601176, 601184, 601178, 601025, 601186, 601110, 601063, 601002, 601029, 601133, 601106, 601215, 601159, 601077, 601202, 601206, 601079, 601051, 601179, 601130, 601058, 601031, 601212, 601173, 601028, 601078, 601094, 601210, 601189, 601228, 601043, 601092, 601171, 601026, 601122, 601208, 601103, 601089, 601108, 601183, 600933 600971, 600919, 600910, 600904, 600935, 601005, 600977, 600936, 600993, 600965, 601011, 600957, 600918, 601015, 600992, 600905, 600914, 600947, 601016, 601013, 600981, 600903, 600949, 601012, 600983, 600970, 600912, 601004, 600915, 600907, 600987, 600939, 600996, 601007, 600899, 601006, 600938, 600952, 600988, 601020, 600928, 600937, 600972, 955056, 965059, 954609, 949471, 954490, 501442, and 499300.
Products Sold
Part Number 973100; Serial Numbers: 600469, 601097, 601098, 601144, 601182, 601160, 601188, 601071, 601117, 601114, 601137, 601059, 601039, 601099, 601088, 601176, 601184, 601178, 601025, 601186, 601110, 601063, 601002, 601029, 601133, 601106, 601215, 601159, 601077, 601202, 601206, 601079, 601051, 601179, 601130, 601058, 601031, 601212, 601173, 601028, 601078, 601094, 601210, 601189, 601228, 601043, 601092, 601171, 601026, 601122, 601208, 601103, 601089, 601108, 601183, 600933 600971, 600919, 600910, 600904, 600935, 601005, 600977, 600936, 600993, 600965, 601011, 600957, 600918, 601015, 600992, 600905, 600914, 600947, 601016, 601013, 600981, 600903, 600949, 601012, 600983, 600970, 600912, 601004, 600915, 600907, 600987, 600939, 600996, 601007, 600899, 601006, 600938, 600952, 600988, 601020, 600928, 600937, 600972, 955056, 965059, 954609, 949471, 954490, 501442, and 499300.
Beckman Coulter Inc is recalling Beckman Coulter UniCel Dxl 800 Access Immunoassay System; Part Number: 973100 With: UniCel Dxl 800 due to Incorrect results: There is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modi. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Incorrect results: There is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range.
Recommended Action
Per FDA guidance
Beckman Coulter mailed a Product Corrective Action (PCA) letter the week of October 8, 2007 to the customers who have the affected UniCel Dxl 800 Access lmmunoassay systems advising that there is an increased variance in assay performance when the UniCel Dxl 800 Modification (MOD ID 10027) and/or UniCel Dxl 800 Modification (MOD ID 6733) have been implemented and there are room fluctuations after assay calibration between the extremes of the 18C and 32C operating range. The letter further advises that Beckman Coulter will contact the customer to arrange a date to remove the modificatlons but until the modification is completed, the system should be operated in a room temperature of 20C to 30C. This can be acomplished by monitoring and recording room tamperature in the area of the instrument. Customers are also asked to review historical quality control records to ensure assay performance is within their laboratory specification and that patient results should be reviewed in context of other diagnostics tests and the clinical presentations of the patient. A response form is included.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026