Beckman Coulter Inc COULTER GEN*S System, Part Number: 6605381, Beckman Coulter, Brea, CA. (Hematology analyzer) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
COULTER GEN*S System, Part Number: 6605381, Beckman Coulter, Brea, CA. (Hematology analyzer)
Brand
Beckman Coulter Inc
Lot Codes / Batch Numbers
Part Number 6605381. All software versions
Products Sold
Part Number 6605381. All software versions
Beckman Coulter Inc is recalling COULTER GEN*S System, Part Number: 6605381, Beckman Coulter, Brea, CA. (Hematology analyzer) due to Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2)- When manually entering a Patient ID, if a blank space is entered between th
Recommended Action
Per FDA guidance
Product Corrective Action (PCA) letters were mailed on April 18,2007, to all GENS, LH500, LH750 and LH780 accounts. They are informed that a potential misidentification can occur in 2 scenarios. 1. When a positive identifier (Sample ID or CassetteIPosition) is manually edited to a positive identifier that is already in the ToDo list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2. When manually entering a Patient ID, if a blank space is entered between the characters in the Patient ID field, the system will only accept the characters before the space.----Short term fix----- The letter explains the issues and provides the customer work around. (Letters were sent by US mail). A customer response form was provided. The root cause of the problem is still under investigation, software upgrades, when approved will be made to devices.