Beckman Coulter Inc GEN*S SlideMaker, Barcode Label Specifications, GEN*S SM Intended Use, GEN*S Slide and Sample, Part Numbers 6605360, P/N 4237221, 4237212, and 4277299. Intended for use as an optional peripheral to the COULTER GEN*S System. Creates a blood smear on a clean microscope slide using a segment of the blood sample aspirated by the GEN*S System. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
GEN*S SlideMaker, Barcode Label Specifications, GEN*S SM Intended Use, GEN*S Slide and Sample, Part Numbers 6605360, P/N 4237221, 4237212, and 4277299. Intended for use as an optional peripheral to the COULTER GEN*S System. Creates a blood smear on a clean microscope slide using a segment of the blood sample aspirated by the GEN*S System.
Brand
Beckman Coulter Inc
Lot Codes / Batch Numbers
All Software Versions.
Products Sold
All Software Versions.
Beckman Coulter Inc is recalling GEN*S SlideMaker, Barcode Label Specifications, GEN*S SM Intended Use, GEN*S Slide and Sample, Part due to Beckman Coulter has become aware of a potential issue in which the Sample ID information embedded in the barcode affixed to a slide made by the Coulte. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Beckman Coulter has become aware of a potential issue in which the Sample ID information embedded in the barcode affixed to a slide made by the Coulter GEN*S and LH SlideMaker does not match the text.
Recommended Action
Per FDA guidance
An Urgent: Product Corrective Action letter, dated June 9, 2008, was sent to customers. The letter identified the affected products, stated the background, discussed the issue, and stated the actions to be taken. Customers are to: 1. Only use upper case alpha characters in the Sample Identifier 2. If you must use lower case alpha characters in the Sample Identifier, configure the SlideMaker Slide labels without a bar-code. Questions regarding the Product Corrective Action should be directed to Beckman Coulter Customer Service at 1-800-526-7694 or your local Beckman Coulter Rep.