Beckman Coulter, Inc. iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325
Brand
Beckman Coulter, Inc.
Lot Codes / Batch Numbers
a) iQ 200 Select, Part Numbers C10684, UDI/DI 15099590695484 and 700-3345, UD/DI 10837461001300, b) iQ 200 Elite, Part Numbers C10683, UDI/DI 15099590697129 and 700-3375, UDI/DI 10837461001355, c) iQ200 Select 1500, Part Number 700-3347, UD/DI 10837461001317, d) iQ200 Sprint, Part Number 700-3325, UDI/DI 10837461001256, All analyzers with Windows 10, Software Version 8.1
Products Sold
a) iQ 200 Select, Part Numbers C10684, UDI/DI 15099590695484 and 700-3345, UD/DI 10837461001300; b) iQ 200 Elite, Part Numbers C10683, UDI/DI 15099590697129 and 700-3375, UDI/DI 10837461001355; c) iQ200 Select 1500, Part Number 700-3347, UD/DI 10837461001317; d) iQ200 Sprint, Part Number 700-3325, UDI/DI 10837461001256; All analyzers with Windows 10, Software Version 8.1;
Beckman Coulter, Inc. is recalling iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 2 due to Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.
Recommended Action
Per FDA guidance
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 04/05/2023 by mail. The notice explained the issue and the hazard and requested the following actions be taken: "Examine your specimen settings to determine if the optional 'Sperm Present' and 'Previous Sample Had Sperm' flags are enabled. If disabled, no further action is needed by your laboratory. If enabled, follow the actions below If the presence of sperm is identified: Review the previous specimen for the presence of sperm. Follow recommendations under Previous Sample Had Sperm in the iQ200 Series Instructions For Use (IFU) 300-4320CE and 300-4321EE and DxU 850m and DxU 840m Iris IFU (C49320AB) in Chapter 6, Data Review, Flags, Previous Sample Had Sperm, Recommendations. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter."