Beckman Coulter Inc Power Processor 3060 Tube Refrigerated Stockyard, P/N 6915556. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Power Processor 3060 Tube Refrigerated Stockyard, P/N 6915556.
Brand
Beckman Coulter Inc
Lot Codes / Batch Numbers
Various.
Products Sold
Various.
Beckman Coulter Inc is recalling Power Processor 3060 Tube Refrigerated Stockyard, P/N 6915556. due to Device does not maintain a uniform internal temperature within 2-8 C at warmer ambient temperatures.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Device does not maintain a uniform internal temperature within 2-8 C at warmer ambient temperatures.
Recommended Action
Per FDA guidance
Recall letters mailed 5/5/2005 request users monitor the temperature variances and evaluate for continued use in thier laboratory.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026