Beckman Coulter Inc. SYNCHRON Ammonia (AMM) Reagent Product Usage: AMM reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems Ammonia Calibrators, is intended for the quantitative determination of Ammonia concentration in human plasma. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
SYNCHRON Ammonia (AMM) Reagent Product Usage: AMM reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems Ammonia Calibrators, is intended for the quantitative determination of Ammonia concentration in human plasma.
Brand
Beckman Coulter Inc.
Lot Codes / Batch Numbers
Lot Number M808278
Products Sold
Lot Number M808278
Beckman Coulter Inc. is recalling SYNCHRON Ammonia (AMM) Reagent Product Usage: AMM reagent, when used in conjunction with SYNCHRON L due to The Ammonia Reagent lot M808278 may fail calibration or cause a low shift in Quality Control recovery on some SYNCHRON Systems.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The Ammonia Reagent lot M808278 may fail calibration or cause a low shift in Quality Control recovery on some SYNCHRON Systems.
Recommended Action
Per FDA guidance
Beckman Coulter mailed a Product Corrective Action (PCA) Letter on April 13, 2009 dated April 10, 2009 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers to discontinue use and discard appropriately the lot number(s) of the affected product and contact their Customer Service Representative at (800) 526-3821 (US), or (800)463-7828 for reagent replacement. Customers were instructed to complete and return the enclosed response form within 10 days so that Beckman Coulter may maintain records. Customers were told to call the Customer Support Center (Hotline) at (800) 854-3633 with any questions regarding the PCA letter.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.