Beckman Coulter Inc Synchron CX4, CX5 and CX9 Pro Systems. Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Synchron CX4, CX5 and CX9 Pro Systems.
Brand
Beckman Coulter Inc
Lot Codes / Batch Numbers
NA
Products Sold
NA
Beckman Coulter Inc is recalling Synchron CX4, CX5 and CX9 Pro Systems. due to Motor driver boards may contain wrong type of microchip which does not have obstruction detection feature.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Motor driver boards may contain wrong type of microchip which does not have obstruction detection feature.
Recommended Action
Per FDA guidance
Product Corrective Action letter was sent the week of 9/29/2003 by US MAIL with a fax back card.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026