Beckman Coulter Inc Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365. Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365.
Brand
Beckman Coulter Inc
Lot Codes / Batch Numbers
Lots: M202040, M206290
Products Sold
Lots: M202040, M206290
Beckman Coulter Inc is recalling Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365. due to The quality control samples for Vancomycin would not recover correctly when compared to the S.I. unit values listed in the incorrectly printed product. Based on FDA medical device recall database.
Summary derived from FDA notice
The following details were not provided by the FDA: distribution information, corrective action details.
Reason for Recall
As stated by FDA
The quality control samples for Vancomycin would not recover correctly when compared to the S.I. unit values listed in the incorrectly printed product insert. The incorrectly printed values were higher than they should be by a factor of 3 (three).
Recommended Action
Per FDA guidance
Healthcare providers and patients should stop using the device and contact the manufacturer for further instructions. Report problems to FDA MedWatch.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026