Becton, Dickinson and Company, BD Biosciences CD11b APC: ASR, REF: 340936, and CE, REF: 333143 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
CD11b APC: ASR, REF: 340936, and CE, REF: 333143
Brand
Becton, Dickinson and Company, BD Biosciences
Lot Codes / Batch Numbers
REF/UDI-DI/Expiration: 340936/00382903409365/20250228, 333143/00382903331437/20250228
Products Sold
REF/UDI-DI/Expiration: 340936/00382903409365/20250228; 333143/00382903331437/20250228
Becton, Dickinson and Company, BD Biosciences is recalling CD11b APC: ASR, REF: 340936, and CE, REF: 333143 due to In vitro diagnostics product that impacts the quality of cell staining can cause low or dim fluorescence signal, which can lead to potential false neg. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
In vitro diagnostics product that impacts the quality of cell staining can cause low or dim fluorescence signal, which can lead to potential false negative results and misdiagnosis of hematological malignancy, delayed treatment, and the need for additional patient samples.
Recommended Action
Per FDA guidance
On 10/25/2023, correction notices were mailed and emailed to customers who were asked to do the following: 1) Discard all products subject to the recall following your institution s process for destruction. 2) Share and post this customer letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness. 3) If lot 3114730 was used, review results and take appropriate action as deemed necessary. 4) Complete and return the response form via email, BDRC26@bd.com or via Link, https://bdx.my.site.com/CC360/s/impactedproducts?rn=BDB-23-4854 Customers who require further assistance can contact the firm's North American Regional Complaint Center via phone, 1-844-823-5433, say "recall" when prompted Mon-Fri 8:00am and 5:00pm CT, or via email, productcomplaints@bd.com
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.