Becton Dickinson & Company PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547 Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547
Brand
Becton Dickinson & Company
Lot Codes / Batch Numbers
UDI-DI: 30382903065470, Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304
Products Sold
UDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304
Becton Dickinson & Company is recalling PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fi due to An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
Recommended Action
Per FDA guidance
An URGENT: Medical Device Product Advisory dated 11/16/23 was sent to customers. Please Take the Following Actions: 1. Immediately review your inventory for the affected product. a. If the product is used for syringe pumps, destroy all affected product subject to this field action following your institution's process for destruction. b. If the product is used for manual application, the product can continue to be used per IFU. 2. If the affected product was previously used on a patient, no further actions are needed. 3. If the affected product is currently in use with a syringe pump, replace the syringe with a non-impacted lot. Please note that healthcare facilities and providers should rely on their syringe pump manufacturer's IFU for the specific syringes that are compatible with their syringe pumps. 4. Share this field action notification with all users of the product within your facility network to ensure they are also aware of this field action. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement if product is used on syringe pumps and has been discarded. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA 1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787 Distributors Please Take the Following Actions: 1. Immediately review your inventory for the affected product. Destroy all unused product subject to the product advisory following your institution s process for destruction. 2. Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026