Becton Dickinson Infusion Therapy Systems Inc. REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO Recall
Hazard assessment based on recall description.
This AI-generated summary is provided for general informational purposes only and is derived from publicly available recall notices. It supplements but does not replace official agency classifications or safety instructions.
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO
Brand
Becton Dickinson Infusion Therapy Systems Inc.
Lot Codes / Batch Numbers
Lot # 2238464/ UDI-DI: (01)00382903868629
Products Sold
Lot # 2238464/ UDI-DI: (01)00382903868629
Becton Dickinson Infusion Therapy Systems Inc. is recalling REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 due to There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect fr. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
Recommended Action
Per FDA guidance
On August 29, 2023, Becton Dickinson issued an "Urgent Medical Device Recall" notification via. FedEx. Becton Dickinson asked consignees to take the following actions: 1. Immediately review your inventory for Catalog Number 386862, Lot number 2238464. Destroy all unused product subject to the recall following your institution's process for destruction. 2. For indwelling catheters, no further action is needed. 3. If a defective product was used on a patient and the used needle was disposed of without the clinician suffering a needlestick injury, no further action is necessary to further protect the patient or the clinician. 4. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch, Phone: 1-800-FDA 1088 (1-800-332-1088), Mail: MedWatch, HF-2, FDA, 5600 Fisher.s Lane, Rockville, MD 20852-978
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.