Bemis Two-Gallon Sharps Container, Models #102 020, #102 030, and #102 040. (The only difference between the models is in color. Model 102 040 is yellow, Model 102 020 is beige and Model 102 030 is red. ) Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Bemis Two-Gallon Sharps Container, Models #102 020, #102 030, and #102 040. (The only difference between the models is in color. Model 102 040 is yellow, Model 102 020 is beige and Model 102 030 is red. )
Lot Codes / Batch Numbers
Lot #20040001 and Lot #20040004
Products Sold
Lot #20040001 and Lot #20040004
A medical device manufacturer is recalling Bemis Two-Gallon Sharps Container, Models #102 020, #102 030, and #102 040. (The only difference be due to The closure flap of the recalled sharps containers have two tabs that fit into slots on the cover for final closure and locking but some of the covers. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The closure flap of the recalled sharps containers have two tabs that fit into slots on the cover for final closure and locking but some of the covers have one of the tabs that will not fit into the designated slot because the slot is occluded with plastic.
Recommended Action
Per FDA guidance
Recall letters dated 2/25/04 requested the return of the product and providing for notification of the subaccounts that may have received the product.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026