Laser Light Show Projectors (Beyond Laser) – Remote Interlock Issue (2025)
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Laser light show projectors manufactured by BLS, Model DIODE 1000/2000/3000.
Brand
Beyond Laser Systems, LLC
Lot Codes / Batch Numbers
N/A
Products Sold
N/A
Beyond Laser Systems, LLC is recalling Laser light show projectors manufactured by BLS, Model DIODE 1000/2000/3000. due to The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically . Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR 1010.4 (Variances), 21 CFR 1010.2 (Certification), 21 CFR 1010.3 (Identification) and 21 CFR 1040.10 (g)(Labelling requirements). The identification label does not include the date of manufacture as required by 21 CFR 1010.3. The corrective action plan covers three families of laser light show of projectors: DIODE. CLUB and PURE with three different hardware configurations such as the number of the number of RJ45 ports, radiant power of emissions and network configurations. No specific user manuals were provided to each model family as required by 21 CFR 1040.10(h)(1) (Informational requirements (1) User information). The Quality Control test did not validate the manual reset function to enable resumption of laser radiation emission from Class IV laser projectors after an interruption caused using a remote interlock. No supplemental product reports, as required according to 21 CFR 1002.11.
Recommended Action
Per FDA guidance
Beyond Laser Systems (BLS) mail the CAP letters to all affected customers. BLS will complete the procedure and ship the laser projector back to the customer within 14-21 days of receiving the units in our facility. BLS will provide a UPS shipping label to the customer for shipping to our facility and cover the expense of shipping the units back to the customer via UPS. If you have any questions, call 512-220-9339 ext 302.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.