Bio-Rad Laboratories Inc Bio-Rad-TOX/See Drug Screen Test (25 tests/box), Catalog number 194-5230; Product is distributed by Bio-Rad Laboratories, Inc., Hercules, CA Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Bio-Rad-TOX/See Drug Screen Test (25 tests/box), Catalog number 194-5230; Product is distributed by Bio-Rad Laboratories, Inc., Hercules, CA
Brand
Bio-Rad Laboratories Inc
Lot Codes / Batch Numbers
DOA6120272
Products Sold
DOA6120272
Bio-Rad Laboratories Inc is recalling Bio-Rad-TOX/See Drug Screen Test (25 tests/box), Catalog number 194-5230; Product is distributed by due to The product may produce a faint line which could be interpreted as a false negative test result .. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
The product may produce a faint line which could be interpreted as a false negative test result .
Recommended Action
Per FDA guidance
The firm intends to contact its subsidiary distributors and sales staff via Fax (Feb 12, 2007) and phone call. In addition, the recall notification will be posted on the firm's intranet site. Customers are instructed to destroy recalled product following standard laboratory practices.