Bio-Rad Laboratories, Inc. BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Kits, REF: 665-1150; and ANA Screen Calibrator Sets, REF: 663-1101 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Kits, REF: 665-1150; and ANA Screen Calibrator Sets, REF: 663-1101
Brand
Bio-Rad Laboratories, Inc.
Lot Codes / Batch Numbers
REF: 663-1131, Lot: 51765, Expiration: 03/05/2021, and any subsequent lots incompatible with: REF: 665-1150, Lot-Expiration: 300884-11/30/2019, 300907-1/31/2020, 300928-3/31/2020, 300933-4/30/2020, and REF: 663-1101, Lot-Expiration: 50661-11/30/2019, 50726-3/11/2020, 50835-3/27/2020, 50858-5/1/2020
Products Sold
REF: 663-1131, Lot: 51765, Expiration: 03/05/2021, and any subsequent lots incompatible with: REF: 665-1150, Lot-Expiration: 300884-11/30/2019, 300907-1/31/2020, 300928-3/31/2020, 300933-4/30/2020; and REF: 663-1101, Lot-Expiration: 50661-11/30/2019, 50726-3/11/2020, 50835-3/27/2020, 50858-5/1/2020
Bio-Rad Laboratories, Inc. is recalling BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Ki due to A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs.
Recommended Action
Per FDA guidance
On 11/01/2019, ANA Screen Lot Matching-Compatibility Notice, were mailed to customers. Customers were advised of the following: If you have received the new control material, please ensure that your laboratory is using one of the newer lot numbers of BioPlex ANA calibrators and ANA reagents. Should you have any of the older BioPlex ANA reagent packs and calibrator sets listed below in ITALICS remaining in your inventory or if you have used these lots in combination with the new control, please contact your local Technical Support center. The following telephone number was provided: 510-724-7000.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.