BioFire Diagnostics, LLC FilmArray Pneumonia Panel (Penumo) REF RFIT-ASY-0144 Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
FilmArray Pneumonia Panel (Penumo) REF RFIT-ASY-0144
Brand
BioFire Diagnostics, LLC
Lot Codes / Batch Numbers
Part Number:RFIT-ASY-0144 UDI: 00815381020178 Kit Lot#: 0938121 Pouch Lot#: 19T621
Products Sold
Part Number:RFIT-ASY-0144 UDI: 00815381020178 Kit Lot#: 0938121 Pouch Lot#: 19T621
BioFire Diagnostics, LLC is recalling FilmArray Pneumonia Panel (Penumo) REF RFIT-ASY-0144 due to Elevated rates of false positive/false negative and control failures while using the Pneumonia Panel, due to issues identified in the manufacturing pr. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Elevated rates of false positive/false negative and control failures while using the Pneumonia Panel, due to issues identified in the manufacturing process.
Recommended Action
Per FDA guidance
On 04/16/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification Letter via email informing its customers of an issue identified in the manufacturing process which might contribute to elevated rates of false positive/false negative results and control failures while using the pneumonia panel. Customer are instructed to: 1) Immediately examine their inventory for product identified in this recall. 2) Discontinue use and discard any remaining product in their possession that is subject to this recall. The Recalling Firm will replace the product at no charge in accordance with its standard limited warranty. 3) Confirm the amount of pouches scrapped (if any) on the attached Acknowledgement of Receipt Form. 4) If the affected products have been further distributed, customers are to identify any recipients and notify them at once of this product recall. 5) Complete the accompanying Field Action Acknowledgment of Receipt Form and return to Recalling Firm so that it may acknowledge their receipt of this notification. For questions or concerns contact Customer Support Department at Support@biofiredx.com or via telephone 1-800-735-6544, option 5 for Product Technical Support.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticePage updated: Jan 10, 2026