Biogenex Laboratories Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B
Brand
Biogenex Laboratories
Lot Codes / Batch Numbers
Serial numbers AS40211, AS40226, AS40230, AS40231, AS40232, AS40233, AS40234, AS40235, AS40236, AS40237. These serial numbers encompass ALL products distributed.
Products Sold
Serial numbers AS40211, AS40226, AS40230, AS40231, AS40232, AS40233, AS40234, AS40235, AS40236, AS40237. These serial numbers encompass ALL products distributed.
Biogenex Laboratories is recalling Xmatrx FISH (fluorescent in situ hybridization) Automated Staining System, Catalog number AS4010B due to Bulk reagent dispensing mechanism may break resulting in reagent spill and possible exposure to chemicals.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Bulk reagent dispensing mechanism may break resulting in reagent spill and possible exposure to chemicals.
Recommended Action
Per FDA guidance
The firm initiated written notification on 01/24/2008. The firm has asked its consignee to inform its consignees of the recall.
Verify this information on the official source