Biomerieux, Inc. NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument). Recall
Check Your Product
According to the U.S. Food and Drug Administration (FDA)
Product
NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).
Brand
Biomerieux, Inc.
Lot Codes / Batch Numbers
Catalog (or Product) number 285204, Serial Number 36061, 36065 and 36068
Products Sold
Catalog (or Product) number 285204, Serial Number 36061, 36065 and 36068
Biomerieux, Inc. is recalling NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instru due to Instrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.. Based on FDA medical device recall database.
Summary derived from FDA notice
Reason for Recall
As stated by FDA
Instrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.
Recommended Action
Per FDA guidance
This correction is being handled as an instrument repair (rework) by bioMerieux Field Service; therefore no communication has been sent to the customer sites. The repair involves heat sink replacement or use of insulating tape on the rear side of the keypad.
Verify this information on the official source
For complete details and official instructions, check the U.S. Food and Drug Administration (FDA) directly.
View official FDA recall noticeDistribution
As reported by FDA
CT, MD, MT